The Escort iMiniPlus PDF mirrors the same specifications as the standard iMiniPlus but can be downloaded without the Console software.  The Logger automatically creates a PDF document upon connection to a PC.  The PDF offers a Graph, Tabular data and a Summary of the data.  If more detailed interrogtaion of the data is required the Logger can still be downloaded using the Console software as normal. 

The iMiniPlus PDF is useful when the monitored product is being shipped to remote sites that may not have the option to install the console software.  The iMiniPlus PDF can also be password protected ensuring security of the recorded data.

Loggers are supplied with USB cables.

  • Measurement range: -40°C to +70°C (-40°F to +158°F)
  • Accuracy:
    • - 40°C to -10°C ±0.5°C (1°F)
    • - 10°C to +70°C ±0.3°C (1°F)
  • Resolution 0.1 °C or 0.1 °F
  • Sensors: Internal & External options
  • Memory: 8048 readings
  • LCD Display
  • Buttons: Start & Stop
  • IP Rating IP65 (with cap over USB connector)
  • Book Marking feature available
  • Size 83 x 57 x 17 mm
  • Software ESCORT Console (or download directly with no software)
  • Download via USB cable, no need for dedicated interface
  • Alarms:
    • Low alarm (consecutive or non-consecutive)
    • High alarm (consecutive or non-consecutive)
    • Also programmable by degree minutes
    • Additional ‘Catastrophe’ alarm available


iMiniPlus PDF Spec sheet

Certification:

Accuracy certification is provided at no cost and 3 point Calibration certification is available upon request.

Supply:

Loggers are shipped pre-configured, ready for use and subject to QC checks before release to document correct configuration against the cutomer order. 

In order to meet demands of short delivery lead times we offer standing orders and inventory maintenance schemes where BioMap holds specific configurations that are replenished when ordered.